Talk Shallowly About OIL-8 Detection Oil Content in Water Detector Machine

Tags: water oil detection

Instrument Function

Both the UV emitting and receiving devices are equipped with an accurate filter system that controls the emission wavelength of the UV light and selects the fluorescence of the specific wavelength that is received by the water sample. The fluorescence intensity of the water sample and the concentration of the oil are in a one-to-one correspondence. After the precise selection of the emission and reception wavelengths by the filtering optical system, the correspondence is more accurate. When cleaning, the sampling plate can be taken out from the sampling tray. Ultraviolet light sources are also easy to replace.

Main Feature

l. UV-fluorescence detection technology

2. Non-contact sensor detection, electrically isolated from the water sample being tested.

3. Continuous online detection and instant reading. (200 measurements per second).

4. The detection area is large, and the sensitivity of measuring the trace oil is high (0.1 ppm).

5. Flexible data processing method, eliminating interference, accurate and reliable data.

6. Standard 4-20mA signal output, visual display of oil concentration values.

7. No need for pharmacy, no consumables, no pollution, and truly environmentally friendly.

8. Automatic or manual multi-point easy calibration, ten groups of storage calibration values are convenient to call.

9. Four relays alarm the measurement overrun and system status.

10. The open sampling chamber structure is easy to maintain.